Additional Information |
|
| Filter Configuration: |
|
1 - HEPA Filter (Certified) 1 - Pre-Filter 2 - UV Lamps |
|
| Product Dimensions: |
|
32" H. x 26" W. x 13" D./Weight- 90lbs. 81.3 cm x 66.0 cm x 33.0 cm - 40.9 kg. |
|
| UV Lamps: |
|
2 |
|
| Applications: |
|
Isolation rooms ICU rooms Examination rooms Interview rooms Small dental operatories Any area where floor and wall space are at a premium |
|
| Electrical Specs: |
|
115V-60Hz |
|
| Amps: |
|
1.75 |
|
| Pressure Gauge: |
|
0-1 inches (0-2.54 cm) |
|
| UV Dosage: |
|
Dosage in Kill Zone - 19,334 uw-sec/cm2 |
|
| Hour Meter: |
|
99,999 hrs. |
|
| Warranty: |
|
2 Years |
|
| Notes: |
|
NQ Environmental Inc. reserves the right to upgrade engineering specifications without notice. |
|
| Decibel Levels: |
|
40 dBA |
|
| Number of Fan Speeds: |
|
1 (250 cfm - may lower it internally) |
|
| Air Delivery (with new filters installed) per Speed: |
|
CFM - 250, CMM - 7.62 |
|
| Maximum Coverage Area: |
|
235 sq ft (based on an 8 ft ceiling) Max. Room Size for 12 ACH with 8' (2.44m) Ceiling - 156 sq. ft. (12' x 13') |
|
| System Efficiency for Particles: |
|
99.97% at 0.3 microns |
|
| Power Requirements: |
|
120 volt - optional 220 volt |
|
| Energy Consumption (Max): |
|
172 watts |
|
| Power Cable: |
|
Grounded - 10 ft. - 'Hospital Grade' / 'Green Dot' (UL498) / (NEMA5-15P) |
|
| Number of Filter Stages: |
|
2 |
|
| Filter Efficiency: |
|
99.97% DOP at .3 microns |
|
| Pre-Filter Efficiency: |
|
40% ASHRAE |
|
| Pre-Filter Frame Material: |
|
V-shaped design, 5 lbs of activated carbon pellets and alumina impregnated with potassium permanganate (50/50 by volume) |
|